airmoVOC C6-C20+

Technical Specifications

airmoVOC C6-C20+ - Volatile and Semi Volatile Hydrocarbons analyzer
  • Detection Limit:
    • Benzene: 5 ppt or 0.0164 µg/m3
    • Naphthalene: 10 ppt or 0.05 µg/m3
  • Detection Range:
    • Benzene: 0.1 to 380 µg/m3
    • Napththalene: 0.05 to 400 µg/m3
  • Relative Standard Deviation:
    • < 0.3 % over 48h (RT)
    • < 3 % over 48 h (Conc.)
  • Results:
    • Data storage
    • Full result storage (data and chromatogram)
    • Embedded computer Windows® based with LCD display
    • 128 GB of Hardware storage on SSD memory
    • 4 USB Connecting Port
    • Two RS-232 ports
    • Display: 10’’ TFT Color LCD
  • Cycle time:
    • 30 min to 60 min (depending on application)
    • 90 min to 120 min (in option)
  • Gas supply:
    • H2 (FID and carrier gas): 30 mL/min (inlet 4 bars; 1/16’’ double ferrule)
    • Air (FID): 180 mL/min (inlet 4 bars ; 1/8’’ double ferrule)
    • Sample inlet (vacuum pump) 1/4’’ double ferrule
    • Pneumatic valve 90ml/ actuation
  • Sample volume:
    • Up to 2200 mL (programmable)

Volatile and Semi Volatile Hydrocarbons analyzer

A GC/FID instrument for Volatile and Semi Volatile Hydrocarbons measurement in ambient air

The airmoVOC C6C20+ is used to monitor VOCs up to the semi volatile hydrocarbons in industrial site, tropical forest and on site under decontamination.
With detection limit as low as 1 ppt for Benzene or Naphthalene the system can measure gaseous PAHs in ambient air in continue up to Fluorene and linear alcanes up to N-Hexadecane when coupled to Mass Spectrometer.

The fully automatic system allows to detect such compounds thanks to internal sampling line and ultra-high sensitivity.

Specific Vistachrom Software is used to pilot the instrument as well as the peripherals (e.g. gas generators) which can be provided with the instrument.