Technical Specifications
- Detection limits:
- H2S: 0.1 ppm (0.1417 mg/m3), mercaptans: 0.1 ppm
- H2S: 5 ppb (7.0 µg/m3), mercaptans: 5 ppb
- Detection range:
- 0/10 or 0/100 or 0/1000 (ppb or ppm)
- Relative Standard Deviation:
- RSD < 3 %: on concentration over 48 h
- RSD < 0.5 %: on retention time over 48 h
- Results:
- Embedded computer Windows® based with LCD display
- 32 GB of Hardware storage on SSD memory
- MODBUS protocol included in standard
- Cycle time:
- H2S/MM/EM : 300s
- H2S, mercaptans1, THT: 720s
- H2Sn, mercaptans1, THT: 900s
- H2S, mercaptants2, THT: 1200s
- Linearity:
- R² > 0.995 for all compounds
- Gas supply:
- Carrier: Dry air or N2 (3 bar): < 4 ml/min
- Use N2 if THT is present
- Internal calibration: 50 ml/min for ppm range
- Sample inlet 1 bar
- Pneumatic valve 90 ml/commutation
H2S, mercaptans, THT, sulphides, total sulfur and odor measurement
Online analysis and monitoring of sulfur compounds in natural gas
Description :
The energyMEDOR is an automatic and isothermal gas
chromatograph dedicated to sulfur compounds analysis (H2S, mercaptans and sulphides) in different matrices.
Two versions exist : the energyMEDOR ppm which measures at ppm levels and the energyMEDOR ppb which measures at ppb levels.
Principle :
- Loop injection by automatic valve
- Detection is made by a gas-liquid reaction
- Sulfur specific detector – no interferences
- The detector life is in excess of 10 years
Features :
- Online, continuous sampling
- No interference from other compounds
- Low maintenance instrument – detector life may be up to 10 years
- Automatic Validation of results by Internal Permeation Tube
- Modifiable threshold alarms – High / Low alarm for odorising processes
VISTACHROM ® software
Chromatotec developed software system enables :
- Remote monitoring & injection control
- Full traceability with on board archiving of results
- QC Set up and control of threshold alarms
- Export of data MODBUS / MGS1 / 4-20mA / 0-10V
TSP Download