H2S MEDOR

Technical Specifications

H2S MEDOR - Hydrogene sulfide
  • Detection limits:
    • H2S : 0.1 ppm (0.1417 mg/m3)
    • H2S : < 0.5 ppm (50µl loop)
  • Detection range:
    • 0/10 or 0/100 or 0/1000 (ppb or ppm)
    • % with internal loop valve
  • Relative Standard Deviation:
    • ≤ 5% on concentration results of H2S STD (24 hours minimum)
    • RSD ≤ 1% and SD < 1s on retention times of H2S standard (24 hours minimum)
  • Results:
    • Embedded computer Windows® based with LCD display
    • 128 GB of Hardware storage on SSD memory
    • MODBUS RTU included in standard
    • Ethernet, remote control
  • Linearity:
    • > 0.995
  • Cycle time:
    • H2S 120 s for 0 – 3 ppm range
    • H2S 300 s for higher range
  • Gas supply:
    • Carrier: Dry air or N2 (3 bar): < 4 ml/min
    • Internal calibration: 50 ml/min for ppm range
    • Sample inlet 1 bar and 160 ml/min sample flow
    • Pneumatic valve 90 ml/commutation

Hydrogene sulfide

Analyzer for continuous on-line monitoring of Hydrogen Sulphide

Description :

H2S Medor is an online Gas Chromatograph dedicated to the analysis of H2S present in different matrices.
Two versions of the instrument exist : 

  • H2S Medor ppm (measuring at ppm levels)
  • H2S Medor ppb (measuring at ppb levels)

Principle :

  • Internal pressure & temperature monitoring
  • Sample loop injection is made by automatic valve
  • H2S Medor uses a sulphur specific electrochemical cell for the detection of H2S

Features :

  • Online, continuous sampling
  • No interference from other compounds
  • Low maintenance instrument – detector life may be up to 10 years
  • Automatic Validation of results by Internal Permeation Tube
  • Modifiable threshold alarms – High / Low alarm for deodorising processes

VISTACHROM ® software 
 Chromatotec developed software system enables :

  • Remote monitoring & injection control
  • Full traceability with on board archiving of results and QC Set up and control of threshold alarms
  • Export of data MODBUS / MGS1 / 4-20mA / 0-10V