airmo HCHO

Technical Specifications

airmoHCHO - Formaldehyde monitoring at low concentration in ambient air or pure gases.
  • Detection Limit:
    • less than 4 ppb for formaldehyde
  • Detection Range:
    • 0 – 100 ppb / amplification 3 in standard
    • 0 – 1000 ppb
    • 0 – 10 000 ppb
  • Relative Standard Deviation:
    • < 0.3 %over 48h (RT)
    • < 2 % over 48 h (Conc.)
  • Results:
    • Data storage
    • 4-20mA current output (option)
    • MODBUS / JBUS or MGS1 communication protocol (option)
  • Cycle time:
    • 15 min as standard (20 min or 30 min optional)
  • Gas supply:
    • H2 (FID and carrier gas): 30 ml/min
    • Air (FID): 180 ml/min
    • N2 (permeation oven)
    • Sample inlet (vacuum pump)
  • Sample volume:
    • 50 ml in standard

Formaldehyde analyzer

airmoHCHO is an autoGC using FID detector dedicated to Formaldehyde monitoring at low concentration in ambient air or pure gases.

This instrument uses a valve with a single sorbent trap, a metallic column and a methanizer. The column is installed in a programmable temperature gradient oven. Formaldehyde reacts with a catalyst for more sensitivity with FID.

Calib is a very important option for good result in ppb range. Permeation tube is used for calibration and data validation.

Miniaturization, sensitivity, mobility and flexibility are its main features.

Everything from the sample port up to the data storage is integrated in a 19″-rack 5U.

Before final delivery, the analyzer is tested for three weeks in the quality control department.

The Vistachrom software enables the user to view and store data on an industrial computer. It provides comfortable user-friendly utilities to recalculate, calibrate and export data and to configure the measurement. The software allows the calculation of retention time, area, mass or concentration profiles.

It can be associated with other Chromatotec’s gas analyzers to enlarge the range of molecules quantified.

TSP Download

Tests reports & Certifications

N/A