Technical Specifications
- Detection Limit:
- Benzene down to 10 ppt with Trap GC/FID/QMS
- Naphthalene: 10 ppt or 0.05 μg/m³
- Detection Range:
- ppt to ppb
- 0.5 to 45 μg/m³ (mCerts tests) with direct injection with QMS: 500 ppb to %
- Relative Standard Deviation (Repeatability):
- < 1% over 48h (RT)
- < 3% over 48h (Concentration)
- Cycle Time:
- 30, 60, 90, or 120 minutes (depending on applications)
- Results:
- Data storage (timestamp)
- 4-20 mA current output (optional)
- MODBUS / JBUS or MGS1 communication protocol (optional)
- Supervisor:
- Embedded computer Windows® based with LED display
- 64 GB hardware storage on SSD memory
- Gas Supply:
- H2 (FID and carrier gas): 45 ml/min (inlet 4 bars; 1/16’’ swagelock)
- Air (FID): 450 ml/min (inlet 3 bars; 1/8’’ swagelock)
- Sample inlet (vacuum pump) 1/8’’ swagelock or by liquid injector
- Power Supply:
- Main (230 V / 115 V, 50 Hz/60 Hz)
- Electrical Consumption:
- 1730 W maximum
- Dimensions:
- 19’’ cabinet: 1850 mm with wheels and lifting rings
- Width: 600 mm
- Depth: 800 mm
- Height: 1850 mm
Transportable Ultra-compact Trap GC-MS / FID system
The Auto GC-MS C10C40+ is a transportable, ultra-compact gas chromatography-mass spectrometry system designed for continuous analysis of light semi-volatile organic compounds (SVOCs) ranging from C10 to C40. This combination of GC and QMS provides a turnkey solution for quantifying and identifying chemicals at low ppt to percentage levels in gases, liquids, and solids.
Key Features:
- Comprehensive Sample Analysis:
- Capable of analyzing gas, liquid, and solid samples, including Bisphenol-A and methyl stearate.
- Validates alkanes up to C40 (boiling point ~522°C).
- Detects PAHs up to C22, including benzo[ghi]perylene (boiling point ~550°C).
- Advanced Technology Integration:
- Equipped with a heated selection valve for GC and an internal cooling system for the mass spectrometer.
- Integrated hydrogen generator (Hydroxychrom USB 4U) and airmoCAL calibration system with permeation tubes.
- Flexible Sampling and Detection:
- Direct air monitoring via injection into the mass spectrometer or using gas chromatography.
- Heated lines for sampling and a flexible injector for liquid or solid analysis.
- Software Control:
- Managed by Vistachrom and VISTAMS software, enabling full traceability, remote monitoring, and injection control.
- Features include archiving of results and chromatograms, setting QC alarms, and data export via MODBUS/4-20 mA / 0-10 V.
Applications:
- Continuous monitoring of SVOCs and PAHs in ambient air, emissions, and industrial environments.
- Analysis of contaminants extracted from solid materials such as plastics (e.g., PET) and industrial oils.
- Quantification of chemicals in gas and liquid samples for environmental monitoring and process control.
Tests reports & Certifications