airmoMEDOR

Technical Specifications

airmoMEDOR - analysis and monitoring of odorous sulfur compounds
  • Detection limits:
    • DES and DMS < 150 ppt (0.15 ppb)
    • DMDS, MM and EM < 50 ppt (0.05 ppb)
    • H2S and SO2 option < 5 ppb
  • Detection range:
    • 0 to 100 ppb
    • Other range on demand
    • Calculation: Odor Index, Total Sulfur, Total mercaptans
  • Relative Standard Deviation:
    • RSD < 3 % on concentration over 48H
    • RSD < 0.5% on retention time over 48H
  • Results:
    • Embedded computer Windows® based with LED display
    • 128 GB of Hardware storage on SSD memory
    • Ethernet, remote control
  • Linearity:
    • 0.995 for all compounds
  • Cycle time:
    • MM, EM, DMS, DMDS, DES 900s
    • With H2S and SO2 option 2400s
    • With IPM, TBM, NPM, 2BM, IBM, NBM and MES option 1800s
  • Gas supply:
    • Carrier: Dry air or N2 (3 bars): ≤ 4 mL / min.
    • CALIB: in continuous ~ 50 mL / min.
    • CALIB during validation ~ 250mL/min
    • Pneumatic valve 90mL/commutation

Online analysis and monitoring of odorous sulfur compounds

The airmoMEDOR® is an automatic gas chromatograph (autoGC) for the analysis and monitoring of trace amounts of Methyl-Mercaptan, Ethyl-Mercaptan, Di-Methyl-Sulfide, Di-Methyl-Di-Sulfide and Di-Ethyl-Sulfide by electrochemical wet cell detection (MEDOR® technology).

The airmoMEDOR® can be used for :

  • monitoring of urban and non-urban pollution
  • monitoring of industrial nuisance
  • deodorization process
  • fugitive emission

TSP Download

Tests reports & Certifications