Technical Specifications

- Configurations and compounds:
- BTEX: Benzene, Cyclohexane, Toluene, Ethylbenzene, m- and p-Xylenes, Styrene and o-Xylene. 1,3-butadiene, IAA or naphthalene are optional.
- C6-C12 PAMS: 33 VOCs from benzene to n-undecane.
- European C6-C12 configuration: 45 VOCs from benzene to n-decane, including biogenic VOCs (terpenes). Other method libraries, such as C3-C10, are optional.
- Detection and linearity:
- Automatic-mode LDL: ≤ 10 ppt. The TSP cites benzene or other VOCs (µ-A76022); performance depends on compound and configuration.
- Benzene ranges: 0-10 ppb, 0-100 ppb or 0-1000 ppb.
- Area linearity: R² > 0.995.
- Repeatability:
- Relative standard deviation over 48 hours: better than 1% for retention time; better than 2% for concentration at 1 ppb.
- Cycle times:
- 10, 15, 20 or 30 minutes (µ-A73022); 30 minutes (µ-A73022-B).
- Sampling and carrier gas:
- Programmable sample volume: 20 to 700 mL or more.
- BTEX and C6-C12 carrier-gas consumption: minimum 3 mL/min with an electric injection valve; varies with column internal diameter and application.
- Internal nitrogen tank: up to 15 barg. Analysis up to 22 hours at 15 bar.
- Detector and column:
- Heated photoionization detector (PID) with a 10.6 eV lamp.
- Electrical six-port valve, single adsorbent trap and capillary column. Programmed column-oven temperature gradient up to 200°C, with carrier-gas pressure/flow control.
- Operating temperature:
- 0°C to 40°C; may vary with the method used.
- Power and battery:
- 12 V DC; battery autonomy up to 8 hours, depending on the method used.
- Battery rechargeable from 230 V / 110 V, 50 Hz / 60 Hz.
- Dimensions and weight:
- Height: 410 mm; width: 340 mm; depth: 180 mm. Net weight: 7 kg.
- Software, storage and communication:
- Vistachrom LITE on Linux; Wi-Fi, Ethernet and one USB port.
- Standard data capacity: 16 GB. Internal data storage up to 10 years; data transfer via USB.
- Measurement configuration, calibration, recalculation and export; retention-time, area, mass and concentration profiles.
- Tablet specifications:
- 8.7-inch display, 1340 × 800 resolution; Android operating system; shock protection included.
- Approximately 14 hours of continuous use and approximately 335 g. Characteristics may vary.
Portable GC-PID VOC Analyzer
µ-airTOXIC is a portable GC-PID analyzer with pre-concentration for trace-level VOC monitoring. Multiple configurations cover BTEX, 1,3-butadiene and VOCs from C3 to C12, according to the selected method.
The C6-C12 versions include a PAMS configuration for 33 VOCs and a European configuration for 45 VOCs, including terpenes. Optional method libraries extend the available applications.
Applications
- Air-quality monitoring in urban areas, environmental monitoring and refinery fence lines.
- Indoor and outdoor air, industrial hygiene and CO₂ beverages.
Operating principle and field deployment
An electrical six-port valve, a single adsorbent trap and a capillary column are combined with a heated PID and a 10.6 eV lamp. Temperature programming of the column oven and carrier-gas pressure/flow control support the selected analytical method.
- The portable suitcase integrates sampling, analysis, data storage, a battery and an internal nitrogen tank, with readiness for operation in less than 5 minutes.
- Integrated Wi-Fi enables control from a phone, tablet or laptop. A local touchscreen is optional.
- Vistachrom LITE provides measurement configuration, calibration, recalculation, data storage and export.
Options
- Portable permeation oven with certified permeation tubes for calibration; nitrogen generator up to 300 mL/min with an external compressor for operation and/or tank refilling, providing up to 10 hours of operation after refilling.
- Modem support and remote control; purge module for extracting VOCs from water for online BTEX analysis; additional battery.
- Rechargeable internal high-pressure carrier-gas cylinder up to 150 barg; external nitrogen bottle for up to 20 days of operation at 70 bar.
- Multiple methods with a VOC library of more than 40 compounds (µ-airTOXIC C6-C12). Biogenic measurement, including isoprene, is available, with potential co-elution of C4 or C5 ozone VOC precursors.
Application-dependent compliance
The TSP states support for compliance with EN 14662-3:2015 for selected applications and lists EN 15267-1, EN 15267-2 and QAL1. Applicability depends on the selected configuration and application; refer to the TSP for details.

TSP Download
Standards, Approvals & Certifications

N/A