Auto GCMS C6C20

Technical Specifications

  • Detection Limit:
    • Benzene down to 10 ppt with Trap GC/FID/QMS
    • Naphthalene: 10 ppt or 0.05 μg/m³
  • Detection Range:
    • ppt to ppb
    • 0.5 to 45 μg/m³ (mCerts tests) with direct injection with QMS: 500 ppb to %
  • Relative Standard Deviation (Repeatability):
    • < 1% over 48h (RT)
    • < 3% over 48h (Concentration)
  • Cycle Time:
    • 30, 60, 90, or 120 minutes (depending on applications)
  • Results:
    • Data storage (timestamp)
    • 4-20 mA current output (optional)
    • MODBUS / JBUS or MGS1 communication protocol (optional)
  • Supervisor:
    • Embedded computer Windows® based with LED display
    • 64 GB hardware storage on SSD memory
  • Gas Supply:
    • H2 (FID and carrier gas): 50 ml/min (inlet 4 bars; 1/16’’ swagelock)
    • Air (FID): 350 ml/min (inlet 3 bars; 1/8’’ swagelock)
    • Sample inlet (vacuum pump) 1/8’’ swagelock or by liquid injector
  • Power Supply:
    • Main (230 V / 115 V, 50 Hz/60 Hz)
  • Electrical Consumption:
    • 1730 W maximum
  • Dimensions:
    • 19’’ cabinet: 1850 mm with wheels and lifting rings
    • Width: 600 mm
    • Depth: 800 mm
    • Height: 1850 mm

Transportable Ultra-compact Trap GC-MS / FID system

The Auto GC-MS C6C20 is a transportable ultra-compact gas chromatography-mass spectrometry system with FID detection, providing a turnkey solution for the continuous analysis of aromatic hydrocarbon compounds (PAH), BTEX, and dioxin precursors from gas, liquid, and solid samples. This combination of GC and QMS allows for the quantification and identification of chemicals at levels from percentage down to low ppt, making it highly versatile for various applications.

Key Features:

  • Comprehensive Sample Analysis:
    • Able to analyze gases, liquids, and solids, including Bisphenol-A, methyl stearate, and other complex compounds.
    • Automatic calculation of toxicity index and total aromatics for VOCs, SVOCs, PAHs, and dioxin precursors.
  • Advanced Technology Integration:
    • Equipped with heated selection valve for GC, mass spectrometer with pumps, and integrated PC with LED display.
    • Built-in hydrogen generator (Hydroxychrom USB 4U) and airmoCAL calibration system with permeation tubes.
  • Flexible Sampling and Detection:
    • Direct air monitoring via injection into the mass spectrometer or gas chromatography.
    • Inbuilt heated line for sampling.
    • Optional features for liquid sample analysis using purge and trap.
  • Software Control:
    • Managed by Vistachrom and MS software, enabling full data traceability, remote monitoring, and injection control.
    • Features include archiving of results and chromatograms, threshold alarm setup, and data export via MODBUS or 4-20 mA / 0-10 V output.

Applications:

  • Ambient air monitoring
  • Cleanroom monitoring
  • PET quality and impurity quantification
  • Contaminant and odorant quantification
  • Process control for hydrocarbons and dioxin precursors

Tests reports & Certifications