Technical Specifications
- Detection Limit:
- Benzene: 5 ppt or 0.0164 µg/m3
- Naphthalene: 10 ppt or 0.05 µg/m3
- Detection Range:
- Benzene: 0.1 to 380 µg/m3
- Napththalene: 0.05 to 400 µg/m3
- Relative Standard Deviation:
- < 0.3 % over 48h (RT)
- < 3 % over 48 h (Conc.)
- Results:
- Data storage
- Full result storage (data and chromatogram)
- Embedded computer Windows® based with LCD display
- 128 GB of Hardware storage on SSD memory
- 4 USB Connecting Port
- Two RS-232 ports
- Display: 10’’ TFT Color LCD
- Cycle time:
- 30 min to 60 min (depending on application)
- 90 min to 120 min (in option)
- Gas supply:
- H2 (FID and carrier gas): 30 mL/min (inlet 4 bars; 1/16’’ double ferrule)
- Air (FID): 180 mL/min (inlet 4 bars ; 1/8’’ double ferrule)
- Sample inlet (vacuum pump) 1/4’’ double ferrule
- Pneumatic valve 90ml/ actuation
- Sample volume:
- Up to 2200 mL (programmable)
Volatile and Semi Volatile Hydrocarbons analyzer
A GC/FID instrument for Volatile and Semi Volatile Hydrocarbons measurement in ambient air
The airmoVOC C6C20 is used to monitor VOCs up to the semi volatile hydrocarbons in industrial site, tropical forest and on site under decontamination.
With detection limit as low as 1 ppt for Benzene or Naphthalene the system can measure gaseous PAHs in ambient air in continue up to Fluorene and linear alcanes up to N-Hexadecane when coupled to Mass Spectrometer.
The fully automatic system allows to detect such compounds thanks to internal sampling line and ultra-high sensitivity.
Specific Vistachrom Software is used to pilot the instrument as well as the peripherals (e.g. gas generators) which can be provided with the instrument.