airmoF

Technical Specifications

airmoF - Automatic Online Formaldehyde analyzer
  • Detection Limit:
    • < 0.1 µg/m³ or < 0.08 ppb
  • Detection Range:
    • 0-400 µg/m³
  • Relative Standard Deviation Precision:
    • Better than 3% over 48h
  • Detection Type:
    • Detector: Fluorescence spectrometry (PhotoMultiplier)
    • Temporal Resolution: 5/h (20 min cycle)
    • Cycle Time: 20 minutes
    • Operating Conditions: Gas T°: 5 – 40°C; Gas RH: 20 – 80%; Atmospheric pressure
    • Calibration: Gaseous Formaldehyde from cylinder or embedded permeation tube
  • Results on LCD:
    • Data storage
    • 4 or 8 x 4-20 mA or 0/10 V output option
    • MODBUS RTU / JBUS or Bayern Hessen protocol
  • Sampling and Gas/Liquid Supply:
    • Sample Flow: 250 ml/min
    • Zero air or nitrogen: 250 ml/min
    • Liquid Reagent Consumption: 30 ml/day
    • Supply Connection: 1/8’’ for gas
  • Instrument Supply:
    • Power Supply: Input 100 – 240V ±10%; 1.5 A max; 47 – 63Hz / Output 15V; 6.67A 100W
    • Power Consumption: Max 75 W
  • General:
    • Rack 19’’ version: Height 222mm, Width 482mm, Depth 600mm, Net weight 15kg
    • Operational Conditions: 0 – 40°C / 20 – 80% RH
    • Storage Conditions: 0 – 40°C / 0 – 85% RH
  • Software & Communication:
    • Expert and standard modes; Data saving; Analysis setting, launching, and monitoring; Defects and maintenance management
    • USB: Data transfer (area, retention time, concentration)
    • Ethernet: Communication and remote control
  • Accessories:
    • DNPH tubes and rings
    • Particle filter, filter strainer
    • 1 L reagent bottle, 1 L waste bottle, 500 mL water bottle with storage and analysis caps with ferrules, bottle holder

Automatic Online Formaldehyde Analyzer

The airmoF is an automatic online analyzer specifically designed for the continuous and real-time monitoring of formaldehyde in various environments. Utilizing microfluidic Hantzsch reaction with fluorescence spectrometry, this device is highly sensitive, accurate, and easy to use.

It is especially well-suited for applications such as public building occupational exposure verification, industrial hygiene measurements, chambers tests studies, material emissions quantification, building management, medical studies, and continuous ambient air monitoring.

Key Benefits:

  • User-Friendly:
    • No gas cylinder needed
    • No filter
    • Low reagent volume & small quantity of air sample required
    • Calibration from embedded formaldehyde permeation tube
    • Compatibility with canisters and FLEC® System
  • Rapid & Accurate Measurements:
    • Online measurements
    • Detection limit down to 0.1 ppb
    • No interferences
    • Gaseous concentration measurement
  • Advanced Analysis Programming & Data Logging:
    • Color touch screen with standard/expert user modes
    • Sequence programming
    • Data logging for quality control
    • Complementary computer software for efficient and easy data processing

Developed in collaboration with CNRS Strasbourg, airmoF represents an innovative approach to formaldehyde monitoring, integrating advanced research with practical applications in the field.

Applications include:

  • Public buildings occupational exposure verification
  • Industrial hygiene measurement
  • Chambers tests studies
  • Material emissions quantification
  • Building management
  • Medical studies
  • Continuous ambient air monitoring

Options:

  • Sampling Teflon line (OD: 1/8’’; L: 150 cm)
  • Spanner (10 – 11 mm)
  • Printed manual
  • 3G module
  • Report’air: Online measurement reporting tool
  • CALIB

Ordering Information:

  • Model: A14022

TSP Download

Tests reports & Certifications