Technical Specifications
- Detection limits:
- ppm Model – 1 mg/m3
- ppb Model – 0.3 mg/m3
- Detection range:
- ppm model – 1 to 50 mg/m3
- ppb model – 0.3 to 3.0mg/ m3(Ranges adjustable depending on Application)
- Relative Standard Deviation:
- RSD < 2 % on concentration over 48H
- RSD < 0.6% on retention time over 48H
- Results:
- Embeded computer Windows® based with LCD display
- 128 GB of Hardware storage on SSD memory
- Linearity:
- 0.995 for all compounds
- Cycle time:
- THT and TBM 180s (for one stream)
- THT and TBM 240s to 300s (for multi stream)
- Gas supply:
- Carrier Gas: N2 (3 bars): 10 ml/min
- Internal calibration: 50 ml/min (if selected)
- Sample inlet 1 bar (160 ml/min)
- Pneumatic valve 90 ml/commutation
Odor measurement (THT)
Online analysis & control of gas odorisation
Description :
THT Medor is an online Gas Chromatograph dedicated to the analysis of THT present in different gases.
Two versions of the instrument exist :
- THT Medor ppm (measuring at ppm levels)
- THT Medor ppb (measuring at ppb levels)
Principle :
- Internal pressure & temperature monitoring
- Sample loop injection is made by automatic valve
- THT Medor uses a sulphur specific electrochemical cell for the detection of THT
Features :
- Online, continuous sampling
- No interference from other compounds
- Low maintenance instrument – detector life may be up to 10 years
- Automatic Validation of results by Internal Permeation Tube
- Modifiable threshold alarms – High / Low alarm for odorising processes
VISTACHROM ® software :
Chromatotec developed software system enables :
- Remote monitoring injection control
- Full traceability with on board archiving of results
- QC Set up and control of threshold alarms
- Export of data MODBUS / MGS1 / 4-20mA / 0-10V