CHROMA DID

Technical Specifications

airmo HCHO - Formaldehyde analyzer
  • Detection Limit:
    • 50 to 100 ppb depending on application and number of products
    • < 50 ppb with special application
  • Detection Range:
    • ppm or ppb
  • Relative Standard Deviation:
    • RSD < 3% on concentrations for 48H
    • RSD < 0.6% on retention times for 48H.
  • Results:
    • Full result storage (data and graphics)
    • Communication protocol MODBUS/ JBUS or MGS1 (option)
    • Output 4-20mA (option)
  • Cycle time:
    • 10 to 15 minutes depending on the application
  • Gas supply:
    • He as carrier gas (inlet 3 bar; 1/8’’ swagelock or 50mL/min maximum)
    • Input calibration at 1 bar: 1/8’’ swagelock for 50mL/min maximum
    • Sample inlet at 1 bar: 1/8’’ swagelock for 20 to 50mL/min maximum
  • Sample volume:
    • Sampling flow rate (20-50mL/min)
    • Injecting sampling loop 250µL to 1ml

Permanent gases analyzer

The continuous or discontinuous mode permanent gases monitor for laboratory or online.

A highly automated system for routine measurement of permanent gases on the range : ppb or ppm.

  • Process : H2 / N2 / O2 / Ar / Kr / Xe / CO* / CO2 / CH4
  • Quality control in pure gas
  • Impurities in Hydrogen
  • Substance tracer
  • Gas input in tubing
  • NH3 in ambiant air
  • CO / *CO2 in ambiant air
  • CO in *C2H4
  • Impurities in CO2

* Can be analysed by chroma CO ( FID methaniser )

Chroma DID is easy to use for ppm and % range.

TSP Download

Tests reports & Certifications

N/A