Technical Specifications
- Compounds analysed:
- H2S, Me-SH, Et-SH, Isopropyl-SH, Npropyl-SH, 2-butylSH, Terbutyl-SH, DMS, DES, DMDS, COS, CS2 …
- Detection limits:
- 4 ppb double Sulphur Compounds like DMDS, CS2 …
- 7 ppb single Sulphur Compounds (H2S DMS …)
- Detection range:
- 4 ppb to 5 ppm (Low Range)
- 100 ppb to 100 ppm (High Range)
- Relative Standard Deviation:
- RSD < 5% on concentration over 24hrs
- RSD < 1% on Retention Time over 24hrs
- Results:
- Full storage of chromatograms & results
- All analysed dates & times stamped individually
- Cycle time:
- 10 minutes in standard
- Shorter cycle time depend on the application
- Gas supply:
- Carrier Gas – Air or N2 3 bar (5mlLmin) 1/8’’
- Sample Inlet -1 Bar or ambient pressure (airmopump)
- Detector – H2 at 2 Bar (70 ml/min) 1/8’’ – Air at 3 bar (80 ml/min) 1/8’’
- Pneumatic valve 90 mL/min commutation
H2S, COS, CS2, SO2, RSH
Automated online analysis & monitoring of Sulphur Compounds: H2S / COS / CS2 / SO2 / RSH
Chroma S is a fully automatic isothermal gas Chromatograph for the analysis & monitoring of sulphur compounds.
Principle :
- Continuous, uninterrupted sampling
- Sample is injected into column for separation
- Controlled oven selects optimum temperature based on application
- detection achieved by a sulphur specific photometric double flame detector
Features :
- The double flame ensures no quenching occurs
- Automatic linearization of compounds
- Identification of compounds is based on their retention time (Elution from the column)
- Automatic data validation as standard with the internal permeation tube
VISTACHROM ® software :
Chromatotec developed software system allows :
- Remote monitoring
- Full traceability through archiving of results and QC
- Set up and control of threshold alarms
- Export of data MODBUS / MGS1 / 4-20mA /0-10V
Tests reports & Certifications
ASTM
US EPA Method TO-15
US EPA Method 16