ChromEnergy DC12

Technical Specifications

airmo HCHO - Formaldehyde analyzer
  • Detection Limit:
    • Down to 10 ppm on Hexane
  • Detection Range:
    • ppm to %
  • Relative Standard Deviation (Repeatability):
    • RSD from 3% to 0.5% over 48h (concentration)
    • RSD < 0.5% over 48h (retention time)
  • Supervisor:
    • Embedded computer Windows® based with LCD display
    • 32 GB of hardware storage on SSD memory
  • Communication:
    • MODBUS protocol
  • Cycle Time:
    • 6 minutes for C6+ / N2 / CO2 / C1 / C2 / C3 / IC4 / NC4 / NeoC5 / IC5 / NC5 / CO / O2
  • Gas Supply:
    • Carrier gas: He (4 bar; 1/16’’): 18 ml/min
    • Calibration inlet: 1 bar; 1/16’’
    • Sample inlet: 1 bar; 1/16’’
    • Pneumatic valve: 90 ml/commutation
  • Power Supply:
    • Main (230 V / 115 V 50 Hz/60Hz)
  • Electric Consumption:
    • Average: 100 VA
    • Peak: 200 VA
  • Dimensions and Weight:
    • Rack: 19’’ (5U)
    • Height: 222 mm
    • Width: 482 mm
    • Depth: 660 mm
    • Net Weight: 22 kg

Continuous Natural Gas Measurement

The chromEnergy DC12 is an industrial gas chromatograph designed for continuous analysis and monitoring of C1 to C6+ compounds in natural gas and gaseous fuels using TCD detection. This system offers automated measurements within the ppm or percentage range, making it ideal for applications in the petrochemical, gas transport, and process industries.

Description:

The chromEnergy DC12 operates with a dual GC setup that includes two TCD detectors within a single instrument, managed by a central supervisor equipped with Vistachrom software. The first GC handles N2, C1, C2, and CO2 using a precolumn and Porapack column, while the second GC manages C6+ and various hydrocarbons through a different column setup. The system features automatic sampling with loop injection and backflush capability for efficient column cleaning.

Key Features:

  • Continuous online monitoring with automatic sampling
  • Low maintenance requirements with full traceability of results and calibration
  • Configurable calorific value unit
  • Robust safety features including pressure safety at the detector outlet
  • Powerful VISTACHROM® software for remote monitoring, injection control, and data export via MODBUS/4-20 mA/0-10 V
  • Intelligence system with adjustable alarm levels and comprehensive data logging

Main Applications:

  • Heating value measurement
  • Trace contaminant monitoring
  • Pipeline integrity
  • Product quality and process control

Available Options:

  • Explosion-proof Exp box
  • ATEX Zone 2 certification: Ex II 3G Ex pz IIC T4
  • CSA International Class 1 Div 2, group B, C & D certification
  • Multiple stream selector (up to 16 streams with one analyzer)
  • Calculation modules for average, statistics, heating value, etc.
  • 4-20 mA / 0-10 V output modules
  • Electric selection valve to reduce air consumption
  • Gas cleaner
  • 24 V DC power supply
  • Carrier gas purificator
  • Flow control with 4-20 mA

Options:

  • Cabinet for ATEX Zone 2
  • Cabinet for CSA Class 1 Division 2

TSP Download

Tests reports & Certifications